Ysi DIQ/S 182 XT-4 System 182 XT-4 User Manual Page 110

  • Download
  • Add to my manuals
  • Print
  • Page
    / 154
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 109
Outputs System 182 XT-4
6 - 18
ba76026e01 01/2012
6.4.4 Limit indicator
Function The characteristic of the limit indicator is laid down in the Limit value
UL, Limit value LL, Hysteresis UL and Hysteresis LL settings. The
fundamentals of the function are described in the introductory chapter
(see section 6.3.2).
In order to set up the Limit indicator function for a relay output, the relay
output must be linked with a sensor (see section 6.2.1).
Settings
Setting Selection/Values Explanation
Limit values UL main variable
LL main variable
UL adjoining var.
LL adjoining var.
Main variable
designates the actual
measured parameter
of the sensor (e.g. pH,
oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).
Limit value UL Within the measuring
range (sensor-
dependent)
Any upper or lower
limiting value
Limit value LL
Hysteresis UL 0 - 10 % of the
measuring range
Hysteresis for upper or
lower limiting value
Hysteresis LL
Behavior at error Open
Close
Unchanged
The relay opens,
closes or remains
unchanged in the case
of system errors or
sensor errors (see
page 6-34).
Action Open
Close
Relay action
Switching delay 0 ... 3600 s The time period for
which a limiting value
must be exceeded
before the relay
operates.
Prevents frequent
switching for
measured values that
are close to the limiting
value.
Page view 109
1 2 ... 105 106 107 108 109 110 111 112 113 114 115 ... 153 154

Comments to this Manuals

No comments