Ysi IFL 700 IQ User Manual

Browse online or download User Manual for Equipment Ysi IFL 700 IQ. YSI IFL 700 IQ User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - IFL 701 IQ

ba76129e01 10/2012IFL 700 IQIFL 701 IQOperating manualIQ SENSOR NET total suspended solids sensorIFL 700 IQ

Page 2 - © 2012 Xylem Inc

Safety instructions IFL 70x IQ2 - 2ba76129e01 10/20122.2 Safe operation2.2.1 Authorized useThe authorized use of the IFL 70x IQ consists of its use as

Page 3 - IFL 70x IQ - Contents

IFL 70x IQ Commissioning3 - 1ba76129e01 10/20123 Commissioning 3.1 IQ SENSOR NET system requirementsSoftware statuses of thecontroller and terminalcom

Page 4 - Contents IFL 70x IQ

Commissioning IFL 70x IQ3 - 2ba76129e01 10/20123.3.2 General installation conditionsFig. 3-1 Ideal installation environmentImmersion depth(= 0.1 m)Wat

Page 5 - 1Overview

IFL 70x IQ Commissioning3 - 3ba76129e01 10/2012Sensor installation The following conditions should always be met: Vertical Sufficient distance from

Page 6 - Overview IFL 70x IQ

Commissioning IFL 70x IQ3 - 4ba76129e01 10/20123.3.5 Short-term interferences due to obstaclesCertain events may for a short period of time affect or

Page 7 - Ultrasonic

IFL 70x IQ Commissioning3 - 5ba76129e01 10/2012NOTE (only IFL 700 IQ)If the sensor is connected to the IQ SENSOR NET, the mechanical wiper may start m

Page 8

Commissioning IFL 70x IQ3 - 6ba76129e01 10/20123.4 Initial commissioningPrerequisite for the display of measured values of the IFL 70x IQ on a termina

Page 9 - 2 Safety instructions

IFL 70x IQ Commissioning3 - 7ba76129e01 10/2012Fig. 3-3 Sample echo profile (sludge blanket height)1 Entered Water depth (shaded)2 Move the cursor alo

Page 10 - 2.2 Safe operation

Commissioning IFL 70x IQ3 - 8ba76129e01 10/2012Fig. 3-4 Measured value display with main and secondary measured value1 Measured value2 Number of echoe

Page 11 - ENSOR NET system requirements

IFL 70x IQ Commissioning3 - 9ba76129e01 10/20123.5 Setting table for the IFL 70x IQ3.5.1 Sensor settings menuCarrying out settings The sensor settings

Page 12 - Commissioning IFL 70x IQ

IFL 70x IQ2ba76129e01 10/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH 45

Page 13 - IFL 70x IQ Commissioning

Commissioning IFL 70x IQ3 - 10ba76129e01 10/2012Setting Selection/values ExplanationMethod The rising side of the echo is evaluated for the measured v

Page 14

IFL 70x IQ Commissioning3 - 11ba76129e01 10/2012Echo selection With this setting, the echo to be evaluated is determined. The echo is automatically id

Page 15 - SACIQ (SW)

Commissioning IFL 70x IQ3 - 12ba76129e01 10/2012Establishing time 0 ... 120 ... 600 sec Filter that ignores (interfering) echoes whose residence time

Page 16 - Temperature

IFL 70x IQ Commissioning3 - 13ba76129e01 10/20123.5.2 Display/Extras menuCarrying out settings Using <>, select the IFL 70x IQ sensor in the m

Page 17

Commissioning IFL 70x IQ3 - 14ba76129e01 10/2012

Page 18 - 1 Measured value

IFL 70x IQ Measuring4 - 1ba76129e01 10/20124 MeasuringFactors affecting themeasured valueThe following factors have an impact on the measured value:

Page 19

Measuring IFL 70x IQ4 - 2ba76129e01 10/2012

Page 20

IFL 70x IQ Maintenance, cleaning, accessories5 - 1ba76129e01 10/20125 Maintenance, cleaning, accessories5.1 General informationWARNINGContact with the

Page 21

Maintenance, cleaning, accessories IFL 70x IQ5 - 2ba76129e01 10/2012We do not recommend unscrewing the sensor from the sensor connection cable when cl

Page 22

IFL 70x IQ Maintenance, cleaning, accessories5 - 3ba76129e01 10/2012CleaningNOTECarefully clean the wiper of the IFL 700 IQ from outside.5.3 Accessori

Page 23

IFL 70x IQ Contents0 - 1ba76129e01 10/2012IFL 70x IQ - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

Page 24

Maintenance, cleaning, accessories IFL 70x IQ5 - 4ba76129e01 10/2012

Page 25 - 4 Measuring

IFL 70x IQ What to do if...6 - 1ba76129e01 10/20126 What to do if...Mechanical damage tothe sensor Display "----"(no valid measuredvalue) C

Page 26 - Measuring IFL 70x IQ

What to do if... IFL 70x IQ6 - 2ba76129e01 10/2012The measured value isnot within the expectedrange Cause RemedyThe Water depth is not set correctly(e

Page 27 - 5.1 General information

IFL 70x IQ What to do if...6 - 3ba76129e01 10/2012Number and position ofthe echoes changingoften Intensity of the existingechoes changing Cause Reme

Page 28 - CCESSORIES). It is always

What to do if... IFL 70x IQ6 - 4ba76129e01 10/2012

Page 29 - 5.3 Accessories

IFL 70x IQ Technical data7 - 1ba76129e01 10/20127 Technical data7.1 Measurement characteristicsMeasuring principle Ultrasound echo measurementMeasurin

Page 30

Technical data IFL 70x IQ7 - 2ba76129e01 10/20127.3 General dataDimensions (in mm)Weight (without sensorconnection cable)Connection technique Connecti

Page 31 - 6 What to do if

IFL 70x IQ Technical data7 - 3ba76129e01 10/20127.4 Electrical data Nominal voltage Max. 24VDC via the IQ SENSOR NET (for more details, see chapter TE

Page 32 - What to do if... IFL 70x IQ

Technical data IFL 70x IQ7 - 4ba76129e01 10/2012

Page 33 - IFL 70x IQ What to do if

IFL 70x IQ Indexes8 - 1ba76129e01 10/20128 Indexes8.1 Explanation of the messagesThis chapter contains a list of all the message codes and related mes

Page 34

Contents IFL 70x IQ0 - 2ba76129e01 10/20127 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1 Measurement characte

Page 35 - 7 Technical data

Indexes IFL 70x IQ8 - 2ba76129e01 10/20128.1.2 Info messagesThe sensor does not generate any info messages.ESC3Cx Sensor defectiveMessage code Message

Page 36 - 7.3 General data

IFL 70x IQ Indexes8 - 3ba76129e01 10/20128.2 Status infoThe status info is a coded piece of information on the current status of a sensor. Each senso

Page 37 - 7.4 Electrical data

Indexes IFL 70x IQ8 - 4ba76129e01 10/2012

Page 38 - Technical data IFL 70x IQ

IFL 70x IQ Contact Information9 - 1ba76129e01 10/20129 Contact Information9.1 Ordering & Technical SupportWhen placing an order please have the fo

Page 39 - 8 Indexes

Contact Information IFL 70x IQ9 - 2ba76129e01 10/2012

Page 41 - 8.2 Status info

1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com

Page 42 - Indexes IFL 70x IQ

IFL 70x IQ Overview1 - 1ba76129e01 10/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig. 1-1

Page 43 - 9 Contact Information

Overview IFL 70x IQ1 - 2ba76129e01 10/20121.2 Structure of the IFL 70x IQ sludge level sensor StructureFig. 1-2 Structure of the sludge level sensor (

Page 44

IFL 70x IQ Overview1 - 3ba76129e01 10/2012Fig. 1-3 Principle of the ultrasonic measurementWiper (cleaning system) The IFL 700 IQ sensor has a mechanic

Page 45

Overview IFL 70x IQ1 - 4ba76129e01 10/2012

Page 46

IFL 70x IQ Safety instructions2 - 1ba76129e01 10/20122 Safety instructions2.1 Safety information2.1.1 Safety information in the operating manualThis o