Ysi IQ SensorNet MIQ/R6 Module User Manual

Browse online or download User Manual for Equipment Ysi IQ SensorNet MIQ/R6 Module. YSI IQ SensorNet MIQ/R6 Module User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - IQ SENSOR NET

ba76031e01 01/2012IQ SENSOR NETMIQ/R6Operating manualPower!OKIQ SENSOR NET six channel relay output module

Page 2 - © 2012 Xylem Inc

Installation MIQ/R63 - 2ba76031e01 01/20123.3 Electrical connections: General instructionsCable glands All electric cables are fed from below via prep

Page 3 - MIQ/R6 - List of contents

MIQ/R6 Installation3 - 3ba76031e01 01/2012General installationinstructionsObserve the following points when attaching connecting wires to the terminal

Page 4 - List of contents MIQ/R6

Installation MIQ/R63 - 4ba76031e01 01/20123.4 Connections to the relay outputsWarningIf external electrical circuits that are subject to the danger of

Page 5 - 1Overview

MIQ/R6 Installation3 - 5ba76031e01 01/2012Connecting lines to theterminal stripFig. 3-1 Terminal strip with the relay connectionsWarningNo free wires

Page 6 - 1.2 Features of the MIQ/R6

Installation MIQ/R63 - 6ba76031e01 01/20127 Close the module.

Page 7 - 2 Safety instructions

MIQ/R6 Settings4 - 1ba76031e01 01/20124 SettingsThe MIQ/R6 module has six relay outputs.Relay outputs operate as openers (normally closed) or closers

Page 8 - 2.1 Authorized use

Settings MIQ/R64 - 2ba76031e01 01/20124.1 Basic information on relay functionsIn this chapter, you will find general basic information concerning the

Page 9 - 3 Installation

MIQ/R6 Settings4 - 3ba76031e01 01/2012Monitoring limitingvalues using one or tworelaysFig. 4-1 Switching points for relays with the function of a limi

Page 10 - Installation MIQ/R6

Settings MIQ/R64 - 4ba76031e01 01/20124.1.3 Proportional outputIn the case of proportional output, a relay switches cyclically on and off in a defined

Page 11 - MIQ/R6 Installation

MIQ/R6 Settings4 - 5ba76031e01 01/2012Output with one relayFig. 4-2 Output with one relayOutput with two relaysFig. 4-3 Output with two relays10100Pro

Page 12

MIQ/R62ba76031e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH 45387

Page 13

Settings MIQ/R64 - 6ba76031e01 01/2012Pulse width output The output of the pulse width is used, e.g. for controlling valves.Pulse-width regulation cha

Page 14 - 7 Close the module

MIQ/R6 Settings4 - 7ba76031e01 01/2012Frequency output Switching frequency output is used, e.g. for controlling dosing pumps.In contrast to the pulse-

Page 15 - 4 Settings

Settings MIQ/R64 - 8ba76031e01 01/2012Characteristic curves Through the selection of the Start value and End value, the proportional output can be ope

Page 16 - Settings MIQ/R6

MIQ/R6 Settings4 - 9ba76031e01 01/2012Positive characteristiccurveThe proportional output range begins above the initial value. If the proportional ra

Page 17 - MIQ/R6 Settings

Settings MIQ/R64 - 10ba76031e01 01/2012Negative characteristiccurveThe proportional output range begins below the initial value. If the proportional r

Page 18

MIQ/R6 Settings4 - 11ba76031e01 01/20124.2 Entering / editing the name of an output For easier identification of the outputs, an individual name can b

Page 19

Settings MIQ/R64 - 12ba76031e01 01/20124.3 Linking the output with a sensor Fig. 4-13 150 - Settings of outputs and links:Link with...NoteOutputs that

Page 20 - Time [s]

MIQ/R6 Settings4 - 13ba76031e01 01/20124.4 Deleting a link with an outputIf a link from a relay output with a sensor is no longer required, the link c

Page 21

Settings MIQ/R64 - 14ba76031e01 01/20124.5 Setting the relay outputsFig. 4-15 150 - Settings of outputs and links1 Call up the measured value display

Page 22

MIQ/R6 Settings4 - 15ba76031e01 01/2012As soon as a function is selected for a relay output, a relay action can be selected (see section 4.5.1).4.5.1

Page 23

MIQ/R6 List of contents0 - 1ba76031e01 01/2012MIQ/R6 - List of contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 Ho

Page 24

Settings MIQ/R64 - 16ba76031e01 01/20124.5.2 System monitoringFunction The System monitoring enables to monitor system errors. In order to set up the

Page 25

MIQ/R6 Settings4 - 17ba76031e01 01/2012In the following cases, the relay does not open despite an invalid measured value: The sensor is being calibra

Page 26

Settings MIQ/R64 - 18ba76031e01 01/2012Generalsensor errors4.5.4 Limit indicatorFunction The characteristic of the limit indicator is laid down in the

Page 27

MIQ/R6 Settings4 - 19ba76031e01 01/2012Behavior at error OpenCloseUnchangedThe relay opens, closes, or remains unchanged in case of system errors or s

Page 28 - 4.5 Setting the relay outputs

Settings MIQ/R64 - 20ba76031e01 01/20124.5.5 Frequency controllerFunction The characteristic of the frequency output is laid down in the Start value,

Page 29

MIQ/R6 Settings4 - 21ba76031e01 01/20124.5.6 Pulse-width outputFunction The characteristic of the pulse width output is laid down in the Start value,

Page 30

Settings MIQ/R64 - 22ba76031e01 01/20124.5.7 CleaningNoteWhen using the MIQ/CHV PLUS valve module, it is best to set the Cleaning function directly at

Page 31

MIQ/R6 Settings4 - 23ba76031e01 01/2012Settings* With short cleaning intervals, the adjustable values for the Cleaning duration and Adjustment time ar

Page 32

Settings MIQ/R64 - 24ba76031e01 01/2012ExampleFig. 4-16 Cleaning cycleSetting ResultReference time (h): Reference time (min):Interval unit:Cleaning in

Page 33

MIQ/R6 Settings4 - 25ba76031e01 01/2012Canceling a cleaningcycleA running cleaning cycle is canceled: Automatically– If the sensor switches to the in

Page 34

List of contents MIQ/R60 - 2ba76031e01 01/20124.7 Behavior of linked outputs . . . . . . . . . . . . . . . . . . 4-294.7.1 Behavior in case of error

Page 35

Settings MIQ/R64 - 26ba76031e01 01/20124.5.8 Sensor-controlledWith the Sensor-controlled function, the relay is controlled by a linked sensor.Require

Page 36

MIQ/R6 Settings4 - 27ba76031e01 01/20124.5.9 Manual controlFunction With the Manual control function, you can test the operativeness of an instrument

Page 37 - ≤ 10 min max. 60 s max. 120 s

Settings MIQ/R64 - 28ba76031e01 01/20124.6 Checking the status of outputsThis function offers a simple overview of the states of all outputs of a MIQ/

Page 38

MIQ/R6 Settings4 - 29ba76031e01 01/20124.7 Behavior of linked outputs4.7.1 Behavior in case of errorFor linked relay outputs, you can determine the be

Page 39

Settings MIQ/R64 - 30ba76031e01 01/20124.7.2 Behavior in non-operative conditionAn output is non-operative when no function is activated for the outpu

Page 40

MIQ/R6 Settings4 - 31ba76031e01 01/20124.8 Maintenance condition of sensorsThe maintenance condition of sensors serves to calibrate, clean, service an

Page 41

Settings MIQ/R64 - 32ba76031e01 01/20124.8.1 Switching on the maintenance condition4.8.2 Switching off the maintenance condition1 Call up the measured

Page 42

MIQ/R6 Maintenance and cleaning5 - 1ba76031e01 01/20125 Maintenance and cleaning5.1 MaintenanceThe MIQ/R6 requires no special maintenance. The general

Page 43

Maintenance and cleaning MIQ/R65 - 2ba76031e01 01/2012

Page 44

MIQ/R6 Technical data6 - 1ba76031e01 01/20126 Technical dataNoteGeneral technical data on MIQ modules are given in the TECHNICAL DATA chapter of the I

Page 45

MIQ/R6 Overview1 - 1ba76031e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig. 1-1 Str

Page 46

Technical data MIQ/R66 - 2ba76031e01 01/2012RelaysOutput Physically separated from the IQ SENSOR NETMax. switching voltage 240 VAC or 24 VDCMax. switc

Page 47 - 5 Maintenance and cleaning

MIQ/R6 Contact Information7 - 1ba76031e01 01/20127 Contact Information7.1 Ordering & Technical SupportWhen placing an order please have the follow

Page 48

Contact Information MIQ/R67 - 2ba76031e01 01/2012

Page 49 - 6 Technical data

MIQ/R6 Indexes8 - 1ba76031e01 01/20128Indexes8.1 Explanation of the messagesIn this chapter you will find a list with all the message codes and corres

Page 50 - Technical data MIQ/R6

Indexes MIQ/R68 - 2ba76031e01 01/2012

Page 52 - Contact Information MIQ/R6

1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com

Page 53 - 8Indexes

Overview MIQ/R61 - 2ba76031e01 01/20121.2 Features of the MIQ/R6General characteristics The MIQ/R6 output module has six relay outputs. You can link r

Page 54 - Indexes MIQ/R6

MIQ/R6 Safety instructions2 - 1ba76031e01 01/20122 Safety instructionsThis operating manual contains special instructions that must be followed during

Page 55

Safety instructions MIQ/R62 - 2ba76031e01 01/20122.1 Authorized useThe authorized use of the MIQ/R6 consists of providing relay outputs in the IQ SENS

Page 56

MIQ/R6 Installation3 - 1ba76031e01 01/20123 Installation3.1 Scope of deliveryThe scope of delivery of the MIQ/R6 is listed in the INSTALLATION chapter