ba76031e01 01/2012IQ SENSOR NETMIQ/R6Operating manualPower!OKIQ SENSOR NET six channel relay output module
Installation MIQ/R63 - 2ba76031e01 01/20123.3 Electrical connections: General instructionsCable glands All electric cables are fed from below via prep
MIQ/R6 Installation3 - 3ba76031e01 01/2012General installationinstructionsObserve the following points when attaching connecting wires to the terminal
Installation MIQ/R63 - 4ba76031e01 01/20123.4 Connections to the relay outputsWarningIf external electrical circuits that are subject to the danger of
MIQ/R6 Installation3 - 5ba76031e01 01/2012Connecting lines to theterminal stripFig. 3-1 Terminal strip with the relay connectionsWarningNo free wires
Installation MIQ/R63 - 6ba76031e01 01/20127 Close the module.
MIQ/R6 Settings4 - 1ba76031e01 01/20124 SettingsThe MIQ/R6 module has six relay outputs.Relay outputs operate as openers (normally closed) or closers
Settings MIQ/R64 - 2ba76031e01 01/20124.1 Basic information on relay functionsIn this chapter, you will find general basic information concerning the
MIQ/R6 Settings4 - 3ba76031e01 01/2012Monitoring limitingvalues using one or tworelaysFig. 4-1 Switching points for relays with the function of a limi
Settings MIQ/R64 - 4ba76031e01 01/20124.1.3 Proportional outputIn the case of proportional output, a relay switches cyclically on and off in a defined
MIQ/R6 Settings4 - 5ba76031e01 01/2012Output with one relayFig. 4-2 Output with one relayOutput with two relaysFig. 4-3 Output with two relays10100Pro
MIQ/R62ba76031e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH 45387
Settings MIQ/R64 - 6ba76031e01 01/2012Pulse width output The output of the pulse width is used, e.g. for controlling valves.Pulse-width regulation cha
MIQ/R6 Settings4 - 7ba76031e01 01/2012Frequency output Switching frequency output is used, e.g. for controlling dosing pumps.In contrast to the pulse-
Settings MIQ/R64 - 8ba76031e01 01/2012Characteristic curves Through the selection of the Start value and End value, the proportional output can be ope
MIQ/R6 Settings4 - 9ba76031e01 01/2012Positive characteristiccurveThe proportional output range begins above the initial value. If the proportional ra
Settings MIQ/R64 - 10ba76031e01 01/2012Negative characteristiccurveThe proportional output range begins below the initial value. If the proportional r
MIQ/R6 Settings4 - 11ba76031e01 01/20124.2 Entering / editing the name of an output For easier identification of the outputs, an individual name can b
Settings MIQ/R64 - 12ba76031e01 01/20124.3 Linking the output with a sensor Fig. 4-13 150 - Settings of outputs and links:Link with...NoteOutputs that
MIQ/R6 Settings4 - 13ba76031e01 01/20124.4 Deleting a link with an outputIf a link from a relay output with a sensor is no longer required, the link c
Settings MIQ/R64 - 14ba76031e01 01/20124.5 Setting the relay outputsFig. 4-15 150 - Settings of outputs and links1 Call up the measured value display
MIQ/R6 Settings4 - 15ba76031e01 01/2012As soon as a function is selected for a relay output, a relay action can be selected (see section 4.5.1).4.5.1
MIQ/R6 List of contents0 - 1ba76031e01 01/2012MIQ/R6 - List of contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 Ho
Settings MIQ/R64 - 16ba76031e01 01/20124.5.2 System monitoringFunction The System monitoring enables to monitor system errors. In order to set up the
MIQ/R6 Settings4 - 17ba76031e01 01/2012In the following cases, the relay does not open despite an invalid measured value: The sensor is being calibra
Settings MIQ/R64 - 18ba76031e01 01/2012Generalsensor errors4.5.4 Limit indicatorFunction The characteristic of the limit indicator is laid down in the
MIQ/R6 Settings4 - 19ba76031e01 01/2012Behavior at error OpenCloseUnchangedThe relay opens, closes, or remains unchanged in case of system errors or s
Settings MIQ/R64 - 20ba76031e01 01/20124.5.5 Frequency controllerFunction The characteristic of the frequency output is laid down in the Start value,
MIQ/R6 Settings4 - 21ba76031e01 01/20124.5.6 Pulse-width outputFunction The characteristic of the pulse width output is laid down in the Start value,
Settings MIQ/R64 - 22ba76031e01 01/20124.5.7 CleaningNoteWhen using the MIQ/CHV PLUS valve module, it is best to set the Cleaning function directly at
MIQ/R6 Settings4 - 23ba76031e01 01/2012Settings* With short cleaning intervals, the adjustable values for the Cleaning duration and Adjustment time ar
Settings MIQ/R64 - 24ba76031e01 01/2012ExampleFig. 4-16 Cleaning cycleSetting ResultReference time (h): Reference time (min):Interval unit:Cleaning in
MIQ/R6 Settings4 - 25ba76031e01 01/2012Canceling a cleaningcycleA running cleaning cycle is canceled: Automatically– If the sensor switches to the in
List of contents MIQ/R60 - 2ba76031e01 01/20124.7 Behavior of linked outputs . . . . . . . . . . . . . . . . . . 4-294.7.1 Behavior in case of error
Settings MIQ/R64 - 26ba76031e01 01/20124.5.8 Sensor-controlledWith the Sensor-controlled function, the relay is controlled by a linked sensor.Require
MIQ/R6 Settings4 - 27ba76031e01 01/20124.5.9 Manual controlFunction With the Manual control function, you can test the operativeness of an instrument
Settings MIQ/R64 - 28ba76031e01 01/20124.6 Checking the status of outputsThis function offers a simple overview of the states of all outputs of a MIQ/
MIQ/R6 Settings4 - 29ba76031e01 01/20124.7 Behavior of linked outputs4.7.1 Behavior in case of errorFor linked relay outputs, you can determine the be
Settings MIQ/R64 - 30ba76031e01 01/20124.7.2 Behavior in non-operative conditionAn output is non-operative when no function is activated for the outpu
MIQ/R6 Settings4 - 31ba76031e01 01/20124.8 Maintenance condition of sensorsThe maintenance condition of sensors serves to calibrate, clean, service an
Settings MIQ/R64 - 32ba76031e01 01/20124.8.1 Switching on the maintenance condition4.8.2 Switching off the maintenance condition1 Call up the measured
MIQ/R6 Maintenance and cleaning5 - 1ba76031e01 01/20125 Maintenance and cleaning5.1 MaintenanceThe MIQ/R6 requires no special maintenance. The general
Maintenance and cleaning MIQ/R65 - 2ba76031e01 01/2012
MIQ/R6 Technical data6 - 1ba76031e01 01/20126 Technical dataNoteGeneral technical data on MIQ modules are given in the TECHNICAL DATA chapter of the I
MIQ/R6 Overview1 - 1ba76031e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig. 1-1 Str
Technical data MIQ/R66 - 2ba76031e01 01/2012RelaysOutput Physically separated from the IQ SENSOR NETMax. switching voltage 240 VAC or 24 VDCMax. switc
MIQ/R6 Contact Information7 - 1ba76031e01 01/20127 Contact Information7.1 Ordering & Technical SupportWhen placing an order please have the follow
Contact Information MIQ/R67 - 2ba76031e01 01/2012
MIQ/R6 Indexes8 - 1ba76031e01 01/20128Indexes8.1 Explanation of the messagesIn this chapter you will find a list with all the message codes and corres
Indexes MIQ/R68 - 2ba76031e01 01/2012
1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com
Overview MIQ/R61 - 2ba76031e01 01/20121.2 Features of the MIQ/R6General characteristics The MIQ/R6 output module has six relay outputs. You can link r
MIQ/R6 Safety instructions2 - 1ba76031e01 01/20122 Safety instructionsThis operating manual contains special instructions that must be followed during
Safety instructions MIQ/R62 - 2ba76031e01 01/20122.1 Authorized useThe authorized use of the MIQ/R6 consists of providing relay outputs in the IQ SENS
MIQ/R6 Installation3 - 1ba76031e01 01/20123 Installation3.1 Scope of deliveryThe scope of delivery of the MIQ/R6 is listed in the INSTALLATION chapter