Summary of Contents

Page 1 - 700 IQ SW

ba76039e01 01/2012VisoTurb® 700 IQVisoTurb® 700 IQ SWOperating manualIQ SENSOR NET turbidity / suspended solids sensorVisoTurb 700 IQVisoTurb 700 IQ S

Page 2 - © 2012 Xylem Inc

Safety instructions VisoTurb® 700 IQ (SW)2 - 2ba76039e01 01/20122.1 Authorized useThe authorized use of the VisoTurb® 700 IQ (SW) comprises its use as

Page 3 - 700 IQ (SW) - Contents

VisoTurb® 700 IQ (SW) Safety instructions2 - 3ba76039e01 01/2012Safe operation If safe operation is no longer possible, the sensor must be taken out o

Page 4 - Contents VisoTurb

Safety instructions VisoTurb® 700 IQ (SW)2 - 4ba76039e01 01/2012

Page 5 - 1Overview

VisoTurb® 700 IQ (SW) Commissioning3 - 1ba76039e01 01/20123 Commissioning 3.1 Scope of delivery VisoTurb® 700 IQ (SW) turbidity/total suspended solid

Page 6 - 700 IQ (SW) turbidity

Commissioning VisoTurb® 700 IQ (SW)3 - 2ba76039e01 01/20123.2.2 Flow directionAs a general rule, the sapphire disc should be positioned clearly agains

Page 7 - 1.4 Features of the VisoTurb

VisoTurb® 700 IQ (SW) Commissioning3 - 3ba76039e01 01/20123.2.4 Sensor orientationThe sensor has a marking (arrow symbol on the shaft or glue dot on c

Page 8 - Overview VisoTurb

Commissioning VisoTurb® 700 IQ (SW)3 - 4ba76039e01 01/20123.2.5 Distances from the ground and wallNoteIn cases of low turbidity (< 100 FNU), the ef

Page 9 - 2 Safety instructions

VisoTurb® 700 IQ (SW) Commissioning3 - 5ba76039e01 01/20123.3 Installation examplesAs a rule, the VisoTurb® 700 IQ (SW) will measure interference-free

Page 10 - 2.1 Authorized use

Commissioning VisoTurb® 700 IQ (SW)3 - 6ba76039e01 01/20123.3.2 Measuring in an open channel (range > 100 FNU)Example: Outlet of a waste watertreat

Page 11

VisoTurb® 700 IQ (SW) Commissioning3 - 7ba76039e01 01/20123.3.3 Measurement in pipelinesNoteIn cases of low turbidity (< 100 FNU), the effect of th

Page 12 - 700 IQ (SW)

VisoTurb® 700 IQ (SW)2ba76039e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Spr

Page 13 - 3 Commissioning

Commissioning VisoTurb® 700 IQ (SW)3 - 8ba76039e01 01/2012Example: 90 ° pipe installationFig. 3-6 Turbidity sensor in the pipe (90 °)The following poi

Page 14 - 0 ° 20 ° 45 °

VisoTurb® 700 IQ (SW) Commissioning3 - 9ba76039e01 01/2012Fig. 3-7 Marking aid3.4 Commissioning / Readiness for measuring3.4.1 Connect the sensorConne

Page 15 - VisoTurb 700 IQ SW

Commissioning VisoTurb® 700 IQ (SW)3 - 10ba76039e01 01/2012NoteDo not suspend the sensor on the sensor connection cable. Use a sensor holder or an arm

Page 16 - Commissioning VisoTurb

VisoTurb® 700 IQ (SW) Commissioning3 - 11ba76039e01 01/20123.4.2 Setting table VisoTurb® 700 IQ (SW)Setting Selection/values ExplanationMeasuring mode

Page 17 - 3.3 Installation examples

Commissioning VisoTurb® 700 IQ (SW)3 - 12ba76039e01 01/2012 AutoRange 0 ... 0.400 ppm 0 ... 4.00 ppm 0 ... 40.0 ppm 0 ... 400 ppm 0 ... 4000 ppm

Page 18

VisoTurb® 700 IQ (SW) Commissioning3 - 13ba76039e01 01/2012Carrying out settings Switch to the main settings menu from the measured value display with

Page 19 - Direction

Commissioning VisoTurb® 700 IQ (SW)3 - 14ba76039e01 01/2012

Page 20

VisoTurb® 700 IQ (SW) Measuring4 - 1ba76039e01 01/20124 MeasuringThe turbidity measurement in aqueous media with the VisoTurb® 700 IQ (SW) is conducte

Page 21 - Marking aid

Measuring VisoTurb® 700 IQ (SW)4 - 2ba76039e01 01/20124.2 Calibration4.2.1 General informationWhy calibrate? The following factors can change with tim

Page 22

VisoTurb® 700 IQ (SW) Measuring4 - 3ba76039e01 01/20124.2.2 Application offsetIn an optimum installation (sufficient distance to the walls, walls made

Page 23 - Setting table VisoTurb

VisoTurb® 700 IQ (SW) Contents0 - 1ba76039e01 01/2012VisoTurb® 700 IQ (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24 - -20.00 ... +20.00

Measuring VisoTurb® 700 IQ (SW)4 - 4ba76039e01 01/2012Fig. 4-1 Ideal measurement environment for the application offset4.2.3 User calibration for meas

Page 25 - 700 IQ (SW) Commissioning

VisoTurb® 700 IQ (SW) Measuring4 - 5ba76039e01 01/2012NoteTo measure the TSS contents, both values of the reference measurement (total suspended solid

Page 26

Measuring VisoTurb® 700 IQ (SW)4 - 6ba76039e01 01/2012NoteThe total suspended solids measurement is even more precise, the better the current status o

Page 27 - 4 Measuring

VisoTurb® 700 IQ (SW) Maintenance, cleaning, accessories5 - 1ba76039e01 01/20125 Maintenance, cleaning, accessories5.1 General informationWARNINGConta

Page 28 - 4.2 Calibration

Maintenance, cleaning, accessories VisoTurb® 700 IQ (SW)5 - 2ba76039e01 01/2012NoteWe do not recommend to unscrew the sensor from the sensor connectio

Page 29 - field (see section 3.4.2)

VisoTurb® 700 IQ (SW) Maintenance, cleaning, accessories5 - 3ba76039e01 01/20125.3 AccessoriesNoteInformation on other IQ SENSOR NET accessories is gi

Page 30 - Marking in

Maintenance, cleaning, accessories VisoTurb® 700 IQ (SW)5 - 4ba76039e01 01/2012

Page 31 - 700 IQ (SW) Measuring

VisoTurb® 700 IQ (SW) What to do if ...6 - 1ba76039e01 01/20126 What to do if ...The sensorautomatically switchesitself on and offperiodically afterco

Page 32 - Measuring VisoTurb

What to do if ... VisoTurb® 700 IQ (SW)6 - 2ba76039e01 01/2012Measured values toohighMeasured value flashesCause Remedy– Gas bubbles in the medium are

Page 33 - 5.1 General information

VisoTurb® 700 IQ (SW) Technical data7 - 1ba76039e01 01/20127 Technical data7.1 Measuring characteristicsMeasuring principle Procedure for measuring sc

Page 34

Contents VisoTurb® 700 IQ (SW)0 - 2ba76039e01 01/20126 What to do if ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17 Technical

Page 35 - 5.3 Accessories

Technical data VisoTurb® 700 IQ (SW)7 - 2ba76039e01 01/2012Accuracy7.2 Application characteristicsAllowedtemperature rangeAllowed pH range of themeasu

Page 36

VisoTurb® 700 IQ (SW) Technical data7 - 3ba76039e01 01/20127.3 General dataDimensionsWeight (without sensorconnection cable)Connection technique Conne

Page 37 - 6 What to do if

Technical data VisoTurb® 700 IQ (SW)7 - 4ba76039e01 01/2012Cleaning system Ultrasound principleAutomatic sensormonitoring(SensCheck function) Recogni

Page 38 - What to do if ... VisoTurb

VisoTurb® 700 IQ (SW) Contact Information8 - 1ba76039e01 01/20128 Contact Information8.1 Ordering & Technical SupportWhen placing an order please

Page 39 - 7 Technical data

Contact Information VisoTurb® 700 IQ (SW)8 - 2ba76039e01 01/2012

Page 40 - Technical data VisoTurb

VisoTurb® 700 IQ (SW) Indexes9 - 1ba76039e01 01/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes and

Page 41 - 7.3 General data

Indexes VisoTurb® 700 IQ (SW)9 - 2ba76039e01 01/20129.1.2 Info messagesEI2341 Operational voltage too low, no operation possible* Check installation a

Page 42 - 7.4 Electrical data

VisoTurb® 700 IQ (SW) Indexes9 - 3ba76039e01 01/20129.2 Status infoThe status info is a piece of coded information about the current state of a sensor

Page 43 - 8 Contact Information

Indexes VisoTurb® 700 IQ (SW)9 - 4ba76039e01 01/2012

Page 45 - 9 Indexes

VisoTurb® 700 IQ (SW) Overview1 - 1ba76039e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating man

Page 46 - Indexes VisoTurb

1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com

Page 47 - 9.2 Status info

Overview VisoTurb® 700 IQ (SW)1 - 2ba76039e01 01/20121.2 Structure of the VisoTurb® 700 IQ (SW) turbidity sensorFig. 1-2 Structure of the turbidity se

Page 48

VisoTurb® 700 IQ (SW) Overview1 - 3ba76039e01 01/20121.4 Features of the VisoTurb® 700 IQ (SW)Turbidity measurementaccording toEN ISO 7027The turbidit

Page 49

Overview VisoTurb® 700 IQ (SW)1 - 4ba76039e01 01/2012

Page 50

VisoTurb® 700 IQ (SW) Safety instructions2 - 1ba76039e01 01/20122 Safety instructionsThis component operating manual contains special instructions tha