OPERATIONS MANUALba76141e03 03/2014MultiLab 4010(P)-1DIGITAL METER FOR IDS SENSORS (PH/ORP/D.O./COND)4010P-14010-14010-1Printermodul4010P-1
Overview MultiLab 4010-110 ba76141e03 03/2014
Index MultiLab 4010-1100 ba76141e03 03/2014Temperature measurementConductivity . . . . . . . . . . . . . . . . . . . . .55O2 . . . . . . . . . . . .
MultiLab 4010-1 Indexba76141e03 03/2014 101
Appendix MultiLab 4010-1102 ba76141e03 03/201419 Appendix19.1 Oxygen solubility tableSolubility of oxygen in mg/L in water exposed to water-saturated
MultiLab 4010-1 Appendixba76141e03 03/2014 10333.0 7.183 6.843 6.518 6.208 5.912 5.63134.0 7.065 6.732 6.415 6.111 5.822 5.54635.0 6.950 6.624 6.314 6
Appendix MultiLab 4010-1104 ba76141e03 03/201419.2 DO% Calibration valuesPressure Altitude Calibration valueInches Hg mm Hg kPa mbar Feet Meters % Sat
MultiLab 4010-1 Contact Information105ba76141e03 03/201420 Contact Information20.1 Ordering & Technical SupportWhen placing an order please have t
Contact Information MultiLab 4010-1106ba76141e03 03/2014
1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com
MultiLab 4010-1 Safetyba76141e03 03/2014 112 Safety2.1 Safety information2.1.1 Safety information in the operating manualThis operating manual provide
Safety MultiLab 4010-112 ba76141e03 03/20142.2 Safe operation2.2.1 Authorized useThe authorized use of the meter consists exclusively of the measureme
MultiLab 4010-1 Commissioningba76141e03 03/2014 133 Commissioning3.1 Scope of delivery Meter MultiLab 4010-1 / MultiLab 4010P-1 4 batteries 1.5 V Mi
Commissioning MultiLab 4010-114 ba76141e03 03/20143.3.1 Inserting the batteriesYou can operate the meter either with normal batteries or with recharge
MultiLab 4010-1 Commissioningba76141e03 03/2014 153.3.2 Connecting the power pack3.3.3 Mounting the standThe stand base can be mounted at the right si
Operation MultiLab 4010-116 ba76141e03 03/20144Operation4.1 General operating principles4.1.1 KeypadIn this operating manual, keys are indicated by br
MultiLab 4010-1 Operationba76141e03 03/2014 174.1.2 DisplayExample (pH):4.1.3 Status information (meter)1 Status information (sensor)2 Measured value3
Operation MultiLab 4010-118 ba76141e03 03/20144.1.4 Instrument connectors4.1.5 Sensor infoYou can display the current sensor data and sensor settings
MultiLab 4010-1 Operationba76141e03 03/2014 194.2 Switching on the meter2. Display further sensor data (settings) with <F1>/[More].1. Switch the
2 ba76141e03 03/2014MultiLab 4010-1For the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, OH
Operation MultiLab 4010-120 ba76141e03 03/20144.3 Switching off the meter4.4 Login with a user nameAfter activation of the user administration by the
MultiLab 4010-1 Operationba76141e03 03/2014 21Changing thepasswordIf the administrator has set up the access with password protection: Forgotten thepa
Operation MultiLab 4010-122 ba76141e03 03/20144.5 Navigation4.5.1 Operating modes4.5.2 Measured value displayIn the measured value display, you can
MultiLab 4010-1 Operationba76141e03 03/2014 23 SettingsSettings are indicated by a colon. The current setting is displayed on the right-hand side. Th
Operation MultiLab 4010-124 ba76141e03 03/20144.5.4 Example 1 on navigation: Setting the language1. Press the <On/Off> key.The measured value di
MultiLab 4010-1 Operationba76141e03 03/2014 255. Select the General submenu with <><>. The current selection is displayed with a frame.6
Operation MultiLab 4010-126 ba76141e03 03/20144.5.5 Example 2 on navigation: Setting the date and timeThe meter has a clock with a date function. The
MultiLab 4010-1 Operationba76141e03 03/2014 276. Change and confirm the setting with <><> and <MENU/ENTER>.The time is set.7. If n
pH value MultiLab 4010-128 ba76141e03 03/20145 pH value5.1 Measuring5.1.1 Measuring the pH valueNOTEWhen connecting a grounded PC/printer, measurement
MultiLab 4010-1 pH valueba76141e03 03/2014 29Criteria for a stablemeasured valueThe Stability control function checks whether the measured values are
MultiLab 4010-1 - ContentsMultiLab 4010-1 Contentsba76141e03 03/2014 31 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH value MultiLab 4010-130 ba76141e03 03/20145.2 pH calibration5.2.1 Why calibrate?pH electrodes age. This changes the zero point (asymmetry) and slop
MultiLab 4010-1 pH valueba76141e03 03/2014 31Continuing with two-point calibration5. Immerse the sensor in the first buffer solution.6. Start the meas
pH value MultiLab 4010-132 ba76141e03 03/2014Continuing withthree- to five-pointcalibration13. Wait for the measurement with stability control to be c
MultiLab 4010-1 pH valueba76141e03 03/2014 335.2.4 Carrying out manual calibration (ConCal)Make sure that in the sensor menu, Buffer menu, the ConCal
pH value MultiLab 4010-134 ba76141e03 03/2014Continuing with two-point calibration7. Wait for the measurement with stability control to be completed o
MultiLab 4010-1 pH valueba76141e03 03/2014 35Continuing withthree- to five-pointcalibration14. Wait for the measurement with stability control to be c
pH value MultiLab 4010-136 ba76141e03 03/20145.2.5 Calibration pointsCalibration can be performed using one to five buffer solutions in any order (sin
MultiLab 4010-1 pH valueba76141e03 03/2014 375.2.6 Calibration dataDisplaying thecalibration dataThe calibration record of the last calibration is to
pH value MultiLab 4010-138 ba76141e03 03/2014CalibrationevaluationAfter calibrating, the meter automatically evaluates the calibration. The zero point
MultiLab 4010-1 pH valueba76141e03 03/2014 395.2.7 Continuous measurement control (CMC function)The Continuous Measurement Control (CMC function) faci
Contents MultiLab 4010-14 ba76141e03 03/20144.5.5 Example 2 on navigation: Setting the date and time . . . 265 pH value. . . . . . . . . . . . . .
pH value MultiLab 4010-140 ba76141e03 03/20145.2.8 QSC function (sensor quality control)General informationon the QSC functionThe QSC function (Qualit
MultiLab 4010-1 pH valueba76141e03 03/2014 41Carry out the normal calibrations for your special measuring range with your usual standard solutions as
pH value MultiLab 4010-142 ba76141e03 03/2014sor evaluation.Carrying out a QSCcontrol calibrationA QSC control calibration can, e.g. be useful if the
MultiLab 4010-1 ORPba76141e03 03/2014 436ORP6.1 Measuring6.1.1 Measuring the ORPNOTEWhen connecting a grounded PC/printer, measurements cannot be perf
ORP MultiLab 4010-144 ba76141e03 03/2014Criteria for a stablemeasured valueThe Stability control function checks whether the measured values are stabl
MultiLab 4010-1 Dissolved oxygenba76141e03 03/2014 457 Dissolved oxygen7.1 Measuring7.1.1 Measuring D.O.Selecting thedisplayedmeasured parameterYou ca
Dissolved oxygen MultiLab 4010-146 ba76141e03 03/2014Stability control(AutoRead)& HOLD functionThe stability control function (AutoRead) continual
MultiLab 4010-1 Dissolved oxygenba76141e03 03/2014 47Criteria for a stablemeasured valueThe Stability control function checks whether the measured val
Dissolved oxygen MultiLab 4010-148 ba76141e03 03/20147.2.4 1-point calibration Calibration in water vapor-saturated air (Cal) and calibration via Comp
MultiLab 4010-1 Dissolved oxygenba76141e03 03/2014 497.2.5 2-point calibrationCalibration in water vapor-saturated air (Cal) and calibration via Compa
MultiLab 4010-1 Contentsba76141e03 03/2014 59.2.1 Settings for ORP measurements . . . . . . . . . . . . . . . . . . 639.3 D.O. measurement settings
Dissolved oxygen MultiLab 4010-150 ba76141e03 03/20145. Start the measurement with <MENU/ENTER>.The measured value is checked for stability (sta
MultiLab 4010-1 Dissolved oxygenba76141e03 03/2014 517.2.6 Calibration dataDisplaying thecalibration recordThe calibration record of the last calibrat
Dissolved oxygen MultiLab 4010-152 ba76141e03 03/2014CalibrationevaluationAfter the calibration, the meter automatically evaluates the current status
MultiLab 4010-1 Conductivityba76141e03 03/2014 538 Conductivity8.1 Measuring8.1.1 Measuring the conductivityNOTEWhen connecting a grounded PC/printer,
Conductivity MultiLab 4010-154 ba76141e03 03/2014ting for automatic Stability control (see section 9.5.3 AUTOMATIC STABILITY CON-TROL, page 69) in the
MultiLab 4010-1 Conductivityba76141e03 03/2014 558.1.2 Measuring the temperatureFor reproducible conductivity measurements, it is essential to measure
Conductivity MultiLab 4010-156 ba76141e03 03/20148.3 Calibration8.3.1 Why calibrate?Aging slightly changes the cell constant, e. g. due to coatings.
MultiLab 4010-1 Conductivityba76141e03 03/2014 578.3.4 Calibration dataThe calibration data can be displayed and then output to the interface.Displayi
Conductivity MultiLab 4010-158 ba76141e03 03/2014CalibrationevaluationAfter calibration, the meter automatically evaluates the current status of the c
MultiLab 4010-1 Settingsba76141e03 03/2014 599 Settings9.1 pH measurement settings9.1.1 Settings for pH measurements Settings The settings are made in
Contents MultiLab 4010-16 ba76141e03 03/201414 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8814.1 pH .
Settings MultiLab 4010-160 ba76141e03 03/20149.1.2 Buffer sets for calibrationYou can use the buffer sets quoted in the table for an automatic calibra
MultiLab 4010-1 Settingsba76141e03 03/2014 619 Merck 5 * 4.0107.00010.00025 °C10 DIN 19267 1.0904.6506.7909.23025 °C11 Mettler Toledo USA * 1.6794.003
Settings MultiLab 4010-162 ba76141e03 03/2014* Brand names or trade names are trademarks of their respective owners pro-tected by law.9.1.3 Calibratio
MultiLab 4010-1 Settingsba76141e03 03/2014 639.2 ORP measurement settings9.2.1 Settings for ORP measurementsThe settings are made in the menu for meas
Settings MultiLab 4010-164 ba76141e03 03/2014Calibration / Calibration data storage / Number of cal. points121-point calibration or 2-point calibratio
MultiLab 4010-1 Settingsba76141e03 03/2014 659.3.2 Enter Sensor cap coefficients9.3.3 DO % Saturation localIrrespective of the height or air pressure,
Settings MultiLab 4010-166 ba76141e03 03/20149.4 Cond measurement settings9.4.1 Settings for IDS conductivity sensors Settings The settings are made i
MultiLab 4010-1 Settingsba76141e03 03/2014 67Setting menu of4320Temp. comp. (TC) / Linear coeff.0.000 ... 2.000 ... 3.000 %/KCoefficient of the linear
Settings MultiLab 4010-168 ba76141e03 03/20149.5 Sensor-independent settings9.5.1 SystemTo open the Storage & config menu, press the <MENU/ENTE
MultiLab 4010-1 Settingsba76141e03 03/2014 699.5.2 Data storageThis menu contains all functions to display, edit and erase stored measured values.9.5.
MultiLab 4010-1 Overviewba76141e03 03/2014 71 Overview1.1 MultiLab 4010-1 meterThe compact, digital precision meter MultiLab 4010-1 enables you to car
Settings MultiLab 4010-170 ba76141e03 03/2014You can also generally switch on the display illumination (see section 9.5.1 SYSTEM, page 68).9.6 ResetYo
MultiLab 4010-1 Settingsba76141e03 03/2014 71ORP The following settings for ORP measurements are reset to the default settings with the Reset function
Settings MultiLab 4010-172 ba76141e03 03/2014The sensor settings are reset under the Reset menu item in the menu for cali-bration and measurement sett
MultiLab 4010-1 Data storageba76141e03 03/2014 7310 Data storageYou can save measured values (datasets) to the data storage: Manual data storage (see
Data storage MultiLab 4010-174 ba76141e03 03/201410.2 Automatic data storage at intervalsThe data storage interval (Interval) determines the time inte
MultiLab 4010-1 Data storageba76141e03 03/2014 75Starting theautomatic datastorage functionTo start the automatic data storage function, select contin
Data storage MultiLab 4010-176 ba76141e03 03/201410.3 Measured value data storage10.3.1 Editing the measured value data storageThe contents of the man
MultiLab 4010-1 Data storageba76141e03 03/2014 77Displayrepresentationof a datasetRepresentation of adataset(USB output)Quitting the display To quit t
Data storage MultiLab 4010-178 ba76141e03 03/2014 ID number Measured value of the connected sensor Measured temperature value of the connected sens
MultiLab 4010-1 Transmitting data (USB interfaces)ba76141e03 03/2014 7911 Transmitting data (USB interfaces) 11.1 Outputting current measurement data1
Overview MultiLab 4010-18 ba76141e03 03/20141.2 MultiLab 4010P-1 meter with integrated printerThe integrated printer of the MultiLab 4010P-1 enables t
Transmitting data (USB interfaces) MultiLab 4010-180 ba76141e03 03/201411.4 Options for data transmission to a PC Via the USB-B interface you can tran
MultiLab 4010-1 Transmitting data (USB interfaces)ba76141e03 03/2014 8111.6 BOD Analyst ProWith the aid of the BOD Analyst Pro software you can admini
Transmitting data (USB interfaces) MultiLab 4010-182 ba76141e03 03/2014
MultiLab 4010-1 Printer (only MultiLab 4010P-1)ba76141e03 03/2014 8312 Printer (only MultiLab 4010P-1)12.1 Commissioning / switching the printer on or
Printer (only MultiLab 4010P-1) MultiLab 4010-184 ba76141e03 03/201412.2 Operation / printingData are only output to the printer if the following cond
MultiLab 4010-1 Printer (only MultiLab 4010P-1)ba76141e03 03/2014 8512.5 What to do if... / printerIntegrated printerdoes not print Printer operating
Maintenance, cleaning, disposal MultiLab 4010-186 ba76141e03 03/201413 Maintenance, cleaning, disposal13.1 Maintenance13.1.1 General maintenance activ
MultiLab 4010-1 Maintenance, cleaning, disposalba76141e03 03/2014 8713.2 CleaningOccasionally wipe the outside of the measuring instrument with a damp
What to do if... MultiLab 4010-188 ba76141e03 03/201414 What to do if...14.1 pHError messageOFL, UFLError message,ErrorMore information and instructio
MultiLab 4010-1 What to do if...ba76141e03 03/2014 89No stable measuredvalue Obviously incorrectmeasured values14.2 Dissolved oxygenError message,OFL,
MultiLab 4010-1 Overviewba76141e03 03/2014 9 have quick-lock couplings with which to fix the sensors to the meter.Sensor data fromIDS sensorsIDS sens
What to do if... MultiLab 4010-190 ba76141e03 03/2014Error message,Error14.3 ConductivityError message,OFL, UFLThe measured value is outside the measu
MultiLab 4010-1 What to do if...ba76141e03 03/2014 91DisplayInstrument does notreact to keystrokeYou want to knowwhich softwareversion is in themeter
Technical data MultiLab 4010-192 ba76141e03 03/201415 Technical data 15.1 Measuring ranges, resolution, accuracy15.2 General dataMeasuring ranges,accu
MultiLab 4010-1 Technical databa76141e03 03/2014 93USB interface (device) Type USB 1.1USB-B (Device), PCBaud rate Adjustable: 1200, 2400, 4800, 9600,
Firmware update MultiLab 4010-194 ba76141e03 03/201416 Firmware update16.1 Firmware update for the meter MultiLab 4010-1You can find available firmwar
MultiLab 4010-1 Firmware updateba76141e03 03/2014 9516.2 Firmware-Update for IDS SensorsWith the "Firmware Update" program and a PC you can
Glossary MultiLab 4010-196 ba76141e03 03/201417 GlossarypH/ORPConductivityAsymmetry see zero pointElectromotive force ofan electrodeThe electromotive
MultiLab 4010-1 Glossaryba76141e03 03/2014 97Dissolved oxygenGeneral informationSalinity The absolute salinity SA of seawater corresponds to the relat
Glossary MultiLab 4010-198 ba76141e03 03/2014Measured parameter The measured parameter is the physical dimension determined by measuring, e. g. pH, co
MultiLab 4010-1 Indexba76141e03 03/2014 9918 IndexAAir calibration beaker . . . . . . . . . . . . . . . . . 47Automatic switch-off function . . . .