ba76006e01 01/2012IQ SENSOR NETSensoLyt® 700 IQ (SW)Operating manualSensoLyt 700 IQ SWSensoLyt 700 IQIQ SENSOR NET pH/ORP sensor

Page 2 - © 2012 Xylem Inc

Safety SensoLyt® 700 IQ (SW)2 - 4ba76006e01 01/2012

Page 3 - 700 IQ (SW) - Contents

SensoLyt® 700 IQ (SW) Commissioning3 - 1ba76006e01 01/20123 Commissioning 3.1 Scope of delivery SensoLyt® 700 IQ (SW) The sensor is fitted with a pr

Page 4 - Contents SensoLyt

Commissioning SensoLyt® 700 IQ (SW)3 - 2ba76006e01 01/2012Connecting the sensorto the sensorconnection cableFig. 3-1 Connecting the sensor1 Take the p

Page 5 - 1Overview

SensoLyt® 700 IQ (SW) Commissioning3 - 3ba76006e01 01/20123.3 Commissioning / Getting the instrument ready for measuringNoteA KCI-filled plastic cap i

Page 6 - 1.2 Structure of the SensoLyt

Commissioning SensoLyt® 700 IQ (SW)3 - 4ba76006e01 01/2012CAUTIONPush the connected combination electrode into the sensor right up to the stop so that

Page 7 - 2 Safety

SensoLyt® 700 IQ (SW) Commissioning3 - 5ba76006e01 01/20127 Screw the protective hood onto the sensor.8 If required, assign a user-defined name to the

Page 8 - 2.1 Authorized use

Commissioning SensoLyt® 700 IQ (SW)3 - 6ba76006e01 01/20123.4 SensoLyt® 700 IQ (SW) setting tableSetting Selection/values ExplanationMeasuring mode 

Page 9 - 700 IQ (SW) Safety

SensoLyt® 700 IQ (SW) Commissioning3 - 7ba76006e01 01/2012Carrying out settings Using s, switch from the measured value display to the main menu of th

Page 10 - Safety SensoLyt

Commissioning SensoLyt® 700 IQ (SW)3 - 8ba76006e01 01/2012

Page 11 - 3 Commissioning

SensoLyt® 700 IQ (SW) Measuring / Operation4 - 1ba76006e01 01/20124 Measuring / Operation4.1 MeasuringWARNINGContact with the sample can lead to dange

Page 12 - Commissioning SensoLyt

SensoLyt® 700 IQ (SW)2ba76006e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Spr

Page 13 - 700 IQ (SW) Commissioning

Measuring / Operation SensoLyt® 700 IQ (SW)4 - 2ba76006e01 01/2012Maintenance condition During calibration the sensor is in the so-called maintenance

Page 14

SensoLyt® 700 IQ (SW) Measuring / Operation4 - 3ba76006e01 01/20124.2.2 Calibration with CAL TEC AUTOMessages displayed during calibration with CAL TE

Page 15

Measuring / Operation SensoLyt® 700 IQ (SW)4 - 4ba76006e01 01/20124.2.3 Calibration with CAL CON 2PMessages displayed during two-point calibration wit

Page 16 - 700 IQ (SW) setting table

SensoLyt® 700 IQ (SW) Measuring / Operation4 - 5ba76006e01 01/20124.2.4 Calibration with CAL CON 1PMessages displayed during the single-point calibrat

Page 17

Measuring / Operation SensoLyt® 700 IQ (SW)4 - 6ba76006e01 01/20124.2.5 Calibration resultCalibration evaluation After calibrating the system automati

Page 18

SensoLyt® 700 IQ (SW) Measuring / Operation4 - 7ba76006e01 01/20124.2.6 Calibration historyCalibration history(available in theIQ SENSOR NETsystem 202

Page 19 - 4 Measuring / Operation

Measuring / Operation SensoLyt® 700 IQ (SW)4 - 8ba76006e01 01/20124.2.7 Reactivation of the last valid calibrationThe SensoLyt® 700 IQ (SW) enables yo

Page 20

SensoLyt® 700 IQ (SW) Maintenance and changing the electrode5 - 1ba76006e01 01/20125 Maintenance and changing the electrode5.1 General maintenance ins

Page 21

Maintenance and changing the electrode SensoLyt® 700 IQ (SW)5 - 2ba76006e01 01/20125.2 Replacing the combination electrodeIf it is necessary to replac

Page 22

SensoLyt® 700 IQ (SW) Maintenance and changing the electrode5 - 3ba76006e01 01/20124 Unscrew the combination combination electrode from the plug head

Page 23

SensoLyt® 700 IQ (SW) Contents0 - 1ba76006e01 01/2012SensoLyt® 700 IQ (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24

Maintenance and changing the electrode SensoLyt® 700 IQ (SW)5 - 4ba76006e01 01/20125.3 DisposalSensor We recommend disposing of the sensor as electron

Page 25

SensoLyt® 700 IQ (SW) Replacement parts and accessories6 - 1ba76006e01 01/20126 Replacement parts and accessories6.1 Combination electrodespH combinat

Page 26

Replacement parts and accessories SensoLyt® 700 IQ (SW)6 - 2ba76006e01 01/2012

Page 27

SensoLyt® 700 IQ (SW) What to do if...7 - 1ba76006e01 01/20127 What to do if...No measured value Measurement does notfunctionSystem cannot becalibrate

Page 28

What to do if... SensoLyt® 700 IQ (SW)7 - 2ba76006e01 01/2012Measurement providesimplausible measuredvaluesCause Remedy– No calibration performed – Ca

Page 29

SensoLyt® 700 IQ (SW) Technical data8 - 1ba76006e01 01/20128 Technical data8.1 Measurement characteristicsMeasuring principle Potentiometric measureme

Page 30 - 5.3 Disposal

Technical data SensoLyt® 700 IQ (SW)8 - 2ba76006e01 01/2012Pressure resistance Max. allowed overpressure (sensor including connection cable):The Senso

Page 31 - 6.2 General accessories

SensoLyt® 700 IQ (SW) Technical data8 - 3ba76006e01 01/20128.3 General dataDimensionsWeight (without sensorconnection cable andcombination electrode)E

Page 32

Technical data SensoLyt® 700 IQ (SW)8 - 4ba76006e01 01/2012Material (Continued)Automatic sensormonitoring(SensCheck function)Function for monitoring g

Page 33 - 7 What to do if

SensoLyt® 700 IQ (SW) Contact Information9 - 1ba76006e01 01/20129 Contact Information9.1 Ordering & Technical SupportWhen placing an order please

Page 34 - What to do if... SensoLyt

Contents SensoLyt® 700 IQ (SW)0 - 2ba76006e01 01/20128 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18.1 Measureme

Page 35 - 8 Technical data

Contact Information SensoLyt® 700 IQ (SW)9 - 2ba76006e01 01/2012

Page 36 - Technical data SensoLyt

SensoLyt® 700 IQ (SW) Indexes10 - 1ba76006e01 01/201210 Indexes10.1 Explanation of the messagesThis chapter contains a list of all the message codes a

Page 37 - 8.3 General data

Indexes SensoLyt® 700 IQ (SW)10 - 2ba76006e01 01/201210.1.2 Info messagesES1311 Component hardware defective* Contact serviceESA311 SensCheck: pH elec

Page 38 - 8.4 Electrical data

SensoLyt® 700 IQ (SW) Indexes10 - 3ba76006e01 01/201210.2 Status infoThe status info is a coded piece of information on the current status of a sensor

Page 39 - 9 Contact Information

Indexes SensoLyt® 700 IQ (SW)10 - 4ba76006e01 01/2012

Page 41 - 10 Indexes

1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com

Page 42 - Indexes SensoLyt

SensoLyt® 700 IQ (SW) Overview1 - 1ba76006e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NET operatingman

Page 43 - 10.2 Status info

Overview SensoLyt® 700 IQ (SW)1 - 2ba76006e01 01/20121.2 Structure of the SensoLyt® 700 IQ (SW)Fig. 1-2 Structure of the pH/ORP sensor (example, Senso

Page 44

SensoLyt® 700 IQ (SW) Safety2 - 1ba76006e01 01/20122 SafetyThis component operating manual contains special instructions that must be followed in the

Page 45

Safety SensoLyt® 700 IQ (SW)2 - 2ba76006e01 01/20122.1 Authorized useThe authorized use of the SensoLyt® 700 IQ (SW) comprises its use as a pH/ORP sen

Page 46

SensoLyt® 700 IQ (SW) Safety2 - 3ba76006e01 01/2012Safe operation If safe operation is no longer possible, the sensor must be taken out of operation a