ba76006e01 01/2012IQ SENSOR NETSensoLyt® 700 IQ (SW)Operating manualSensoLyt 700 IQ SWSensoLyt 700 IQIQ SENSOR NET pH/ORP sensor
Safety SensoLyt® 700 IQ (SW)2 - 4ba76006e01 01/2012
SensoLyt® 700 IQ (SW) Commissioning3 - 1ba76006e01 01/20123 Commissioning 3.1 Scope of delivery SensoLyt® 700 IQ (SW) The sensor is fitted with a pr
Commissioning SensoLyt® 700 IQ (SW)3 - 2ba76006e01 01/2012Connecting the sensorto the sensorconnection cableFig. 3-1 Connecting the sensor1 Take the p
SensoLyt® 700 IQ (SW) Commissioning3 - 3ba76006e01 01/20123.3 Commissioning / Getting the instrument ready for measuringNoteA KCI-filled plastic cap i
Commissioning SensoLyt® 700 IQ (SW)3 - 4ba76006e01 01/2012CAUTIONPush the connected combination electrode into the sensor right up to the stop so that
SensoLyt® 700 IQ (SW) Commissioning3 - 5ba76006e01 01/20127 Screw the protective hood onto the sensor.8 If required, assign a user-defined name to the
Commissioning SensoLyt® 700 IQ (SW)3 - 6ba76006e01 01/20123.4 SensoLyt® 700 IQ (SW) setting tableSetting Selection/values ExplanationMeasuring mode
SensoLyt® 700 IQ (SW) Commissioning3 - 7ba76006e01 01/2012Carrying out settings Using s, switch from the measured value display to the main menu of th
Commissioning SensoLyt® 700 IQ (SW)3 - 8ba76006e01 01/2012
SensoLyt® 700 IQ (SW) Measuring / Operation4 - 1ba76006e01 01/20124 Measuring / Operation4.1 MeasuringWARNINGContact with the sample can lead to dange
SensoLyt® 700 IQ (SW)2ba76006e01 01/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Spr
Measuring / Operation SensoLyt® 700 IQ (SW)4 - 2ba76006e01 01/2012Maintenance condition During calibration the sensor is in the so-called maintenance
SensoLyt® 700 IQ (SW) Measuring / Operation4 - 3ba76006e01 01/20124.2.2 Calibration with CAL TEC AUTOMessages displayed during calibration with CAL TE
Measuring / Operation SensoLyt® 700 IQ (SW)4 - 4ba76006e01 01/20124.2.3 Calibration with CAL CON 2PMessages displayed during two-point calibration wit
SensoLyt® 700 IQ (SW) Measuring / Operation4 - 5ba76006e01 01/20124.2.4 Calibration with CAL CON 1PMessages displayed during the single-point calibrat
Measuring / Operation SensoLyt® 700 IQ (SW)4 - 6ba76006e01 01/20124.2.5 Calibration resultCalibration evaluation After calibrating the system automati
SensoLyt® 700 IQ (SW) Measuring / Operation4 - 7ba76006e01 01/20124.2.6 Calibration historyCalibration history(available in theIQ SENSOR NETsystem 202
Measuring / Operation SensoLyt® 700 IQ (SW)4 - 8ba76006e01 01/20124.2.7 Reactivation of the last valid calibrationThe SensoLyt® 700 IQ (SW) enables yo
SensoLyt® 700 IQ (SW) Maintenance and changing the electrode5 - 1ba76006e01 01/20125 Maintenance and changing the electrode5.1 General maintenance ins
Maintenance and changing the electrode SensoLyt® 700 IQ (SW)5 - 2ba76006e01 01/20125.2 Replacing the combination electrodeIf it is necessary to replac
SensoLyt® 700 IQ (SW) Maintenance and changing the electrode5 - 3ba76006e01 01/20124 Unscrew the combination combination electrode from the plug head
SensoLyt® 700 IQ (SW) Contents0 - 1ba76006e01 01/2012SensoLyt® 700 IQ (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance and changing the electrode SensoLyt® 700 IQ (SW)5 - 4ba76006e01 01/20125.3 DisposalSensor We recommend disposing of the sensor as electron
SensoLyt® 700 IQ (SW) Replacement parts and accessories6 - 1ba76006e01 01/20126 Replacement parts and accessories6.1 Combination electrodespH combinat
Replacement parts and accessories SensoLyt® 700 IQ (SW)6 - 2ba76006e01 01/2012
SensoLyt® 700 IQ (SW) What to do if...7 - 1ba76006e01 01/20127 What to do if...No measured value Measurement does notfunctionSystem cannot becalibrate
What to do if... SensoLyt® 700 IQ (SW)7 - 2ba76006e01 01/2012Measurement providesimplausible measuredvaluesCause Remedy– No calibration performed – Ca
SensoLyt® 700 IQ (SW) Technical data8 - 1ba76006e01 01/20128 Technical data8.1 Measurement characteristicsMeasuring principle Potentiometric measureme
Technical data SensoLyt® 700 IQ (SW)8 - 2ba76006e01 01/2012Pressure resistance Max. allowed overpressure (sensor including connection cable):The Senso
SensoLyt® 700 IQ (SW) Technical data8 - 3ba76006e01 01/20128.3 General dataDimensionsWeight (without sensorconnection cable andcombination electrode)E
Technical data SensoLyt® 700 IQ (SW)8 - 4ba76006e01 01/2012Material (Continued)Automatic sensormonitoring(SensCheck function)Function for monitoring g
SensoLyt® 700 IQ (SW) Contact Information9 - 1ba76006e01 01/20129 Contact Information9.1 Ordering & Technical SupportWhen placing an order please
Contents SensoLyt® 700 IQ (SW)0 - 2ba76006e01 01/20128 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18.1 Measureme
Contact Information SensoLyt® 700 IQ (SW)9 - 2ba76006e01 01/2012
SensoLyt® 700 IQ (SW) Indexes10 - 1ba76006e01 01/201210 Indexes10.1 Explanation of the messagesThis chapter contains a list of all the message codes a
Indexes SensoLyt® 700 IQ (SW)10 - 2ba76006e01 01/201210.1.2 Info messagesES1311 Component hardware defective* Contact serviceESA311 SensCheck: pH elec
SensoLyt® 700 IQ (SW) Indexes10 - 3ba76006e01 01/201210.2 Status infoThe status info is a coded piece of information on the current status of a sensor
Indexes SensoLyt® 700 IQ (SW)10 - 4ba76006e01 01/2012
1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com
SensoLyt® 700 IQ (SW) Overview1 - 1ba76006e01 01/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NET operatingman
Overview SensoLyt® 700 IQ (SW)1 - 2ba76006e01 01/20121.2 Structure of the SensoLyt® 700 IQ (SW)Fig. 1-2 Structure of the pH/ORP sensor (example, Senso
SensoLyt® 700 IQ (SW) Safety2 - 1ba76006e01 01/20122 SafetyThis component operating manual contains special instructions that must be followed in the
Safety SensoLyt® 700 IQ (SW)2 - 2ba76006e01 01/20122.1 Authorized useThe authorized use of the SensoLyt® 700 IQ (SW) comprises its use as a pH/ORP sen
SensoLyt® 700 IQ (SW) Safety2 - 3ba76006e01 01/2012Safe operation If safe operation is no longer possible, the sensor must be taken out of operation a