OPERATING MANUALba76040e04 05/2014ViSolid®700 IQViSolid®700 IQ SWViSolid®700 IQ HIQ SENSORNET TOTAL SUSPENDED SOLIDS SENSORViSolid 700 IQViSolid 700 I
2 - 2 ba76040e04 05/2014Safety ViSolid®700 IQ (SW)2.2 Safe operation2.2.1 Authorized useThe authorized use of the ViSolid®700 IQ (SW) consists of its
ViSolid®700 IQ (SW) Safetyba76040e04 05/2014 2 - 32.3 Hazardous location use (ViSolid®700 IQ H)Rated sensor models The ViSolid®700 IQ H (YSI-Order No.
2 - 4 ba76040e04 05/2014Safety ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 13 Commissioning 3.1 IQ SENSORNET system requirementsSoftware statuses of thecontroller and te
3 - 2 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)3.3.2 General informationThe measuring principle of the ViSolid®700 IQ (SW) (scattered light
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 33.3.3 Flow directionGenerally, in flowing media, the measurement window should be clearly pit
3 - 4 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)3.3.5 Sensor orientationThe sensor has a marking (arrow symbol on the shaft or glue dot on co
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 53.3.6 Distances from the ground and wallThe following graphic indicates the minimum distances
3 - 6 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)3.4 Installation examplesAs a rule, the ViSolid®700 IQ (SW) will measure interference-free wh
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 7Measurement in achannelIn an open channel, the sensor can be immersed in the sample using a w
ba76040e04 05/2014ViSolid®700 IQ (SW)For the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs, O
3 - 8 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)3.4.2 Measurement in pipelinesIf deposits occur on the pipe walls, the calibration should be
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 9Example: 90 ° pipe installationFig. 3-6 Total suspended solids sensor in a pipe (90 °)The fol
3 - 10 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)Fig. 3-7 Marking aid3.5 Commissioning / Readiness for measuring3.5.1 Connect the sensorConne
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 11them dry using compressed air).Connecting the sensorto the sensorconnection cableFig. 3-8 Co
3 - 12 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)3.5.2 Selecting the Measuring modeSpecify the following data in the Measuring mode setting
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 133.5.3 Setting tableViSolid®700 IQ (SW)Setting Selection/values ExplanationMeasuring mode (se
3 - 14 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)TSS measuring mode:Menu selection:User calibration Basic settings Value pairs 1..3 Value
ViSolid®700 IQ (SW) Commissioningba76040e04 05/2014 3 - 15Carrying out settings Switch to the main settings menu from the measured value display with
3 - 16 ba76040e04 05/2014Commissioning ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Measuringba76040e04 05/2014 4 - 14 MeasuringThe ViSolid®700 IQ (SW) measures the light scattered and reflected by the total suspen
ViSolid®700 IQ (SW) - ContentsViSolid®700 IQ (SW) Contentsba76040e04 05/2014 0 - 11 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 - 2 ba76040e04 05/2014Measuring ViSolid®700 IQ (SW)4.2 Calibration for TSS measurement4.2.1 General informationWhy calibrate? The following factors
ViSolid®700 IQ (SW) Measuringba76040e04 05/2014 4 - 34.2.2 Default calibrationDefault calibration formatrix type 1The factory calibration curve for ma
4 - 4 ba76040e04 05/2014Measuring ViSolid®700 IQ (SW)Default calibration formatrix type 2The factory calibration curve for matrix type 2 was determine
ViSolid®700 IQ (SW) Measuringba76040e04 05/2014 4 - 54.2.3 Correction factorThe setting of the Correction factor provides a simple option for adapting
4 - 6 ba76040e04 05/2014Measuring ViSolid®700 IQ (SW)Determining theCorrection factorSetting the Correctionfactor1 Bring the sensor into the measuring
ViSolid®700 IQ (SW) Measuringba76040e04 05/2014 4 - 74.2.4 User calibration The displayed values of total suspended solids are calculated with the aid
4 - 8 ba76040e04 05/2014Measuring ViSolid®700 IQ (SW)Procedure of thecalibrationA measurement of the total suspended solids will deliver ever more acc
ViSolid®700 IQ (SW) Measuringba76040e04 05/2014 4 - 9Entering the CalibrationdataBelow the smallest value, the calibration curve is extended to the ze
4 - 10 ba76040e04 05/2014Measuring ViSolid®700 IQ (SW)The calibration data entered are evaluated by the system. The entry of the calibration value pai
ViSolid®700 IQ (SW) Maintenance, cleaning, accessoriesba76040e04 05/2014 5 - 15 Maintenance, cleaning, accessories5.1 General informationThe ViSolid®7
0 - 2 ba76040e04 05/2014Contents ViSolid®700 IQ (SW)4.2.2 Default calibration . . . . . . . . . . . . . . . . . . . . 4-34.2.3 Correction factor . .
5 - 2 ba76040e04 05/2014Maintenance, cleaning, accessories ViSolid®700 IQ (SW)If you need to disconnect the sensor from the sensor connection cable, p
ViSolid®700 IQ (SW) Maintenance, cleaning, accessoriesba76040e04 05/2014 5 - 35.3 AccessoriesDescription Model Order no.Screwable plug for sensor conn
5 - 4 ba76040e04 05/2014Maintenance, cleaning, accessories ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) What to do if ...ba76040e04 05/2014 6 - 16 What to do if ...Mechanical damage tothe sensor Display always shows"0" TSS
6 - 2 ba76040e04 05/2014What to do if ... ViSolid®700 IQ (SW)Measured values toolow Measured values toohigh – Signal averaging time too short for lo
ViSolid®700 IQ (SW) Technical databa76040e04 05/2014 7 - 17 Technical data7.1 Measuring characteristicsMeasuring principle Procedure for measuring sca
7 - 2 ba76040e04 05/2014Technical data ViSolid®700 IQ (SW)7.2 Application characteristicsAllowedtemperature rangeAllowed pH range of themeasuring medi
ViSolid®700 IQ (SW) Technical databa76040e04 05/2014 7 - 37.3 General dataDimensions (in mm)Weight (without sensorconnection cable)Connection techniqu
7 - 4 ba76040e04 05/2014Technical data ViSolid®700 IQ (SW)* Stainless steel can be susceptible to corrosion at chloride concentra-tions of ≥ 500 mg/l
ViSolid®700 IQ (SW) Technical databa76040e04 05/2014 7 - 57.5 Electrical data Rated voltage(non-H models)24 V DC via the IQ SENSORNET (for more detail
ViSolid®700 IQ (SW) Overviewba76040e04 05/2014 1 - 11Overview1.1 How to use this component operating manualStructure of theIQ SENSORNET operatingmanua
7 - 6 ba76040e04 05/2014Technical data ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Indexesba76040e04 05/2014 8 - 18 Indexes8.1 Explanation of the messagesThis chapter contains a list of all the message codes and r
8 - 2 ba76040e04 05/2014Indexes ViSolid®700 IQ (SW)8.1.2 Info messagesEC2342 User calibration error, check TSS/SiO2 pairs of variates* All TSS values
ViSolid®700 IQ (SW) Indexesba76040e04 05/2014 8 - 38.2 Status infoThe status info is a piece of coded information about the current state of a sensor.
8 - 4 ba76040e04 05/2014Indexes ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Contact Informationba76040e04 05/2014 9 - 19 Contact Information9.1 Ordering & Technical SupportWhen placing an order please h
9 - 2 ba76040e04 05/2014Contact Information ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Appendixba76040e04 05/2014 10 - 110 Appendix10.1 Check calibration valuesBy checking the value pairs, possible calibration errors
10 - 2 ba76040e04 05/2014Appendix ViSolid®700 IQ (SW)
1 - 2 ba76040e04 05/2014Overview ViSolid®700 IQ (SW)1.2 Structure of the ViSolid®700 IQ (SW) total suspended solids sensorFig. 1-2 Structure of the to
1) The tissue in plants that brings water upward from the roots;2) a leading global water technology company.We're 12,500 people unified in a com
ViSolid®700 IQ (SW) Overviewba76040e04 05/2014 1 - 31.4 Features of the ViSolid®700 IQ (SW)Total suspended solidsmeasurementThe measurement of the tot
1 - 4 ba76040e04 05/2014Overview ViSolid®700 IQ (SW)
ViSolid®700 IQ (SW) Safetyba76040e04 05/2014 2 - 12 Safety2.1 Safety information2.1.1 Hazard warnings in this operating manualThe hazard warnings are